Abstract:Objective:To detect some of biological agents by a newly developed fiber optic biosensor FOB-3.Methods:Yersinia pestis F1 antigen,Bacillus anthraticis vegetative form and its spore and staphylococcal enterotoxin B(SEB)were detected respectively by the FOB-3 with sandwich immunoassay after a series of optimization of the experimental parameters.In order to decrease noise among fluorescent signals,cutoff value was calculated,and results was showed in the form of(Ssignal-Nnoise).Results and Conclusion:F1 antigen at 50~1 000 ng/mL,SEB at 0.1~100 μg/mL,vegetative B.anthraticis at 3×101~3×106 CFU/mL and anthrax spore at 4×104~4×107 CFU/mL were detected with obvious positive signals in 20 min by the FOB-3,which preliminarily realized the advantage of sensitivity,simple-operation in rapid detection of biological agents.
What problem does this paper attempt to address?