Application of canonical correlation analysis (CCA) to correct NIR spectra measured on different instruments and in different measurement dates

Kaiyi Zheng,Wei Fan,Ting Wu,Huilian Hu,Yiping Du,Yizeng Liang
DOI: https://doi.org/10.11719/com.app.chem20130305
2013-01-01
Abstract:Calibration transfer methods were used to correct near infrared spectra (NIR) of tri-component mixtures including methylbenzene, chlorobenzene and n-heptane for their variations resulting from differences of measurement date and instrument. For correcting spectra between different measurement dates and instruments the spectra of mixtures measured in spectrometer of Bruker on date 16 were assigned as primary spectra while the spectra of same mixtures detected in spectrometer of Bruker on date 17 and in spectrometer of Perkin-Elmer on date 16 were set as secondary spectra for changes of measurement date and instrument respectively. After spectra obtained, two calibration transfer methods including canonical correlation analysis (CCA) and piecewise direct standardization (PDS) were used to correct the secondary spectra. The results showed that although variations of both measurement date and instrument can cause secondary spectra to deviate from primary spectra, the CCA can correct the secondary spectra to use the model of primary spectra to predict those three components with high accuracy.