Optimal Design for Open MRI Superconducting Magnet with Active Shielding

Chun-zhong WANG,Qiu-liang WANG,Lan-kai LI,Ming RONG,You-yuan ZHOU,王春忠,王秋良,李兰凯,荣明,周又元
2010-01-01
Abstract:The optimal design method for an open Magnetic Resonance Imaging(MRI)superconducting magnet with an active shielding configuration is proposed.Firstly,three pairs of current rings are employed as seed coils.By optimizing the homogeneity of Diameter Sphere Volume(DSV),the positions and currents of the seed coils will be obtained.Secondly,according to the positions and currents of the seed coils,the current density of superconducting wires is determined,and then the original sections for the coils can be achieved.An optimization for the homogeneity based on the constrained nonlinear optimization method is employed to determine the coils with good homogeneity.Thirdly,the magnetic field generated by previous coils is set as the background field,then add two coils with reverse current,and optimize the stray field line of 5 Gauss in a certain scope.Finally,a further optimization for the homogeneity is used to get final coils.This method can also be used in the design of other axisymmetric superconducting MRI magnets.
What problem does this paper attempt to address?