Analysis of Selective Inspections on Radiation Protection of Medical Diagnostic X-Ray Rooms in Shanghai,2007-2011

陈飚,陈春晖,朱素蓉,周艳琴,朱国英
DOI: https://doi.org/10.13213/j.cnki.jeom.2013.08.015
2013-01-01
Abstract:[Objective]To evaluate the current radiation protection level and trend of medical diagnostic X-ray rooms in Shanghai.[Methods]Selective inspection data of radiation protection of medical diagnostic X-ray rooms between 2007 and 2011 in Shanghai were collected and analyzed based on municipal radiological protection supervision conducting in the same period of time.[Results]From 2007 to 2011,the radiation protection of medical X-ray diagnostic rooms was at a high level,and the qualified rates ranged from 98.8%-100.0%.Medical X-ray CT rooms and radiography rooms were the main room types where unqualified sampling points were identified;and unqualified sampling locations included doors of waiting rooms,armed walls,doors of control rooms,and viewing windows.[Conclusion]The radiation protection of medical diagnostic X-ray rooms is currently at a high level in Shanghai.Regular supervision is an effective way to locate and correct the problems of radiation protection in medical institutions.
What problem does this paper attempt to address?