Improving the Sensitivity of NIR Spectroscopy with an Enrichment Technique: Determining a Trace Analyte of Ethyl Carbamate.

Wei Xuemei,Huang Zixia,Zhang Wei,Du Yiping
DOI: https://doi.org/10.2116/analsci.23.853
IF: 1.967
2007-01-01
Analytical Sciences
Abstract:An exploration was made to develop a determination method of a low-concentration analyte by NIR spectroscopy. An absorber, silica gel was employed to extract and enrich a low-concentration analyte of ethyl carbamate. The solid absorber with the enriched analyte was measured by NIR spectroscopy in the range of 800 - 2500 nm. Afterwards, PLS regression was performed between the NIR spectra and the concentrations of the analyte for quantitative analysis of the low-concentration analyte. The spectra of 20 solid samples of analyte-absorbed silica gel showed a good correlation with the concentrations of ethyl carbamate in the samples. A leave-one-out cross validation was applied to evaluate the prediction ability of PLS models built with the full spectra, spectra in the region of 1920 - 1970 nm and the region of 2250 - 2430 nm, respectively. The values of the root-mean-square error of the cross validation (RMSECV) were about 0.1 mg L(-1) (0.1 ppm).
What problem does this paper attempt to address?