The influence of different media environments on the accuracy of Root ZX measuring the length of root canal in vitro

Chen Yan,Yang Minghua,Zhu Min
DOI: https://doi.org/10.3969/j.issn.1004-7565.2006.02.003
2006-01-01
Abstract:Objective To study the influence of different media environment on the accuracy of Root ZX measuring the length of root canal in vitro.Methods To select 30 teeth separated from body(50 root canal)and the agar model,in 3% H_2O_2 solution,0.9%NaCl solution,0.12% Chlorhexidine solution,17%EDTA solution or dry root canal,measuring the length of root canal using Root ZX was compared with the practice length of root canal.Results In five different environments,there was no statistically significant difference between the measuring length using Root ZX and the real length of root canal(P>0.05);Accurate rate of electronic ohmmeter was 90.4% compared with practice length.Conclusions The accuracy of measure of root ZX was not influenced by root canal media and had no relation to the dry or moist root canal.
What problem does this paper attempt to address?