Observations of oxygen therapy in premature infants

ZHU Jian-hong,YU Ren-jie,WANG Jun-yi,YANG Lin
DOI: https://doi.org/10.3969/j.issn.1673-6710.2008.01.004
2008-01-01
Abstract:Objective Measuring inhaled oxygen concentration from various ways of administration,its flow, environment in order to find an optimal way of oxygen therapy for the premature infants. Methods To measure and compare the oxygen concentrations in different ways of administration of inhaled oxygen including bilateral nasal tube, head hood, Fastar head hood and incubator tent with various oxygen flow. Results In the group using bilateral nasal tube, oxygen concentration was greater than 95% in common method group and 60% in improved method group in which oxygen flow velocity was 0.5 L/min,1 L/min and 2 L/min.The difference showed statistical significance (P<0.05). In the head hood group, measuring oxygen concentration was less than 50% in the 3-5 L/min of oxygen flow rate. Fastar head hood showed the same results. In the incubator tent, measuring oxygen concentration was less than 30% in the 3-5 L/min of oxygen flow rate. Conclusions Head hood with less 5 L/min of oxygen flow velocity and Fastar head hood and incubator tent are the safe ways of oxygen therapy for premature infants.
What problem does this paper attempt to address?