Precision Analyzing on Virtual Volume by Using Respiration Motion Phantom

DAI Wei,DENG Xiao-wu,HUANG Shao-min,HUANG Xiao-yan,ZHANG Li
DOI: https://doi.org/10.3969/j.issn.1005-202x.2007.06.001
2007-01-01
Abstract:Objective: A QUASAR respiration motion phantom was used to study the aliasing of target volume encountered when acquiring treatment planning CT's of a moving target in lung. Materials and Methods: The unit was programmed for (S/I) displacements of ± 1 cm and ± 2 cm respectively, about apoint,with initial phase 0 or π with period of 3 s, 4 s, 5 s, 6 s, 10 s. Targets consisted of two spheres of unit density, 1 cm and 2 cm in diameters and a cube,3 cm in side length. Spiral CT scanning was performed on CT scanner (Siemens) on both stationary and dynamic targets. A pitch of 1.5 was used to acquire CT scan slices of thicknesses 5 mm. Scanspeeds of 1 s were investigated, and a consistent image acquisition to motion phase synchronization scheme was used on all scans that involved moving targets. Result: (1) While moving, there are obviously volume aliasing on S/I direction. Geometric miss, which is predominantly observed at the S/I extremes of the moving target image; (2)The size of target and the motion state both effect Volume aliasing. The max difference of volume reach to 66.7%. While amplitude is 2 cm, target volumes between 2 cm will partly overlapped; (3) Geometric centre of the target volume also changed. Conclusion: When speacial setup, we ought to limit the motion of the patient if his organ amplitude reach to 2 cm and his respiratory cycle is above and beyond 4 s.With the real amplitude of the motion state which directly effect the speacial integrity of target volume. Or evaluating the discrepancy of the plan which based on aliasing of target volume with measured result.
What problem does this paper attempt to address?