The design of 3-D navigator for siting acetabular component and evaluate its accuracy and feasibility

XU Jie,MA Ruo-fan,DING Yue,LIU Shang-li
DOI: https://doi.org/10.3969/j.issn.1674-134X.2008.06.013
2008-01-01
Abstract:Objective To investigate the effectivity and accuracy of 3-D navigator for siting acetabular component without the affect of body position.Methods Site the cup in the pelvis with the traditional mechanical inserter and new 3-D navigator respectively. Conimeter and 3-D image reconstruction with spiral CT were used to measure the abduction and anteversion angles.Results Difference between the setting abduction and anteversion angle of acetabular cup and the real ones were no more than 5° with the help of 3-D navigator. No significant difference of both abduction and anteversion angle for reproducibility of siting cup was found (T test,P>0.05).Conclusions The 3-D navigator is more accurate than the traditional mechanical cup inserter. The duplicate test demonstrates that the repeatability is better for the new device.