[Simultaneous Detection of Enterovirus 71 and Coxsackievirus A16 by Multiplex Real-Time PCR with an Internal Control].

Xinglong Xiao,Yaqing He,Yigang Yu,Hong Yang,Huifang Li,Xiaoquan Yang,Jingwei Zhang,Gu Chen,Dongmei Liu,Xiaofeng Li,Hui Wu
DOI: https://doi.org/10.3321/j.issn:0001-6209.2009.01.017
2009-01-01
Abstract:OBJECTIVE:To rapidly identify EV71 and CA16 simultaneously.METHODS:A multiplex real-time PCR with an internal control (IC) was developed. The specificity, sensitivity, reproducibility of the real time RT-PCR assay were estimated and more over 400 clinical samples were tested.RESULTS:The results showed 100% specificity for the selected panel. The assay met the sensitivity of 50% tissue culture infective dose (TCID50) per milliliter samples for CA16 and 0.1 TCID50 for EV71. Analysis with 10(4) - 0.1 TCID50/mL EV71 or CA16 samples demonstrated high reproducibility with a coefficient of variation (CV) of 0.9% - 2.0% for EV71 and 0.9% - 2.3% for CA16. More than 400 clinical samples were detected by real-time PCR,The results showed that the average positive ratio for EV71 and CA16 were 46.1% and 14.2%, higher than common assay (34.5% and 12.8%). Moreover, the result statistics indicated that there were PCR inhibition in stools, rectal swabs and throat swabs specimens with inhibition ratio from 1.8% to 3.4%. The inhibition ratio of these samples showed the importance of IC when PCR was used to detect the RNA of EV71, CA16 or other enterovirus.CONCLUSION:As a result of its high specificity, sensitivity and avoiding false negative results by using an internal control, the assay is suitable for rapid clinical diagnosis of EV71 and CA16.