Performance Of Tiloronoxim And Tilorone Determination In Human Blood By Hplc-Ms/Ms: Method Validation, Uncertainty Assessment And Its Application To A Pharmacokinetic Study

Xianhua Zhang,Jingli Duan,Suodi Zhai,Yiheng Yang,Li Yang
DOI: https://doi.org/10.1016/j.jchromb.2009.12.016
2010-01-01
Abstract:A highly sensitive and selective HPLC-MS/MS method is presented for the quantitative determination of tiloronoxim and its metabolite tilorone in human blood. An aliquot of 200 mu l human blood was extracted with a mixture of chloroform/ethyl ether(1/2, v/v), using metoprolol as the internal standard (the IS). Separation was achieved on an Xterra MS C18 column (50 mm x 2.1 mm, 5 mu m) with a gradient mobile phase of methanol/water containing 15 mM ammonium bicarbonate (pH 10.5). Detection was performed using positive MRM mode on a TurbolonSpray source. The mass transitions monitored were m/z 426.3 -> 100.0, m/z 411.3 -> 100.0 and m/z 2683 -> 116.1 for tiloronoxim, tilorone and the IS, respectively. The method was fully validated using total error theory, which is based on beta-expectation tolerance intervals and include trueness and intermediate precision. The method was found to be accurate over a concentration range of 1-100 ng/ml for both compounds. The measurement uncertainty based on P-expectation tolerance intervals was assessed at each concentration level of the validation standards. This method was successively applied to a pharmacokinetic study of tiloronoxim in healthy volunteers. (C) 2009 Elsevier B.V. All rights reserved