Oxygenated hemoglobin diffuse reflectance ratio for in vitro detection of human gastric pre-cancer

L. Q. Li,H. J. Wei,Z. Y. Guo,H. Q. Yang,G. Y. Wu,S. S. Xie,H. Q. Zhong,X. Y. Li,Q. L. Zhao,X. Guo
DOI: https://doi.org/10.1134/S1054660X10130116
IF: 1.2
2010-01-01
Laser Physics
Abstract:Oxygenated hemoglobin diffuse reflectance (DR) ratio (R540/R575) method based on DR spectral signatures is used for early diagnosis of malignant lesions of human gastric epithelial tissues in vitro. The DR spectra for four different kinds of gastric epithelial tissues were measured using a spectrometer with an integrating sphere detector in the spectral range from 400 to 650 nm. The results of measurement showed that the average DR spectral intensity for the epithelial tissues of normal stomach is higher than that for the epithelial tissues of chronic and malignant stomach and that for the epithelial tissues of chronic gastric ulcer is higher than that for the epithelial tissues of malignant stomach. The average DR spectra for four different kinds of gastric epithelial tissues show dips at 542 and 577 nm owing to absorption from oxygenated Hemoglobin (HbO 2 ). The differences in the mean R540/R575 ratios of HbO 2 bands are 6.84% between the epithelial tissues of normal stomach and chronic gastric ulcer, 14.7% between the epithelial tissues of normal stomach and poorly differentiated gastric adenocarcinoma and 22.6% between the epithelial tissues of normal stomach and undifferentiated gastric adenocarcinoma. It is evident from results that there were significant differences in the mean R540/R575 ratios of HbO 2 bands for four different kinds of gastric epithelial tissues in vitro ( P < 0.01).