Temperature Measurement of Cold Atoms in A Cesium Magneto-Optical Trap by Means of Short-Distance Time-Of-Flight Absorption Spectrum

T Geng,SB Yan,YH Wang,HJ Yang,TC Zhang,JM Wang
DOI: https://doi.org/10.7498/aps.54.5104
IF: 0.906
2005-01-01
Acta Physica Sinica
Abstract:We report the basic idea and experimental demonstration of measuring the temperature of cold cesium atomic cloud confined in a magneto-optical trap (MOT) by analyzing absorption spectrum observed in the short-distance time-of-flight (TOF) method. Compared with the conventional TOF scheme, we measure the temperature of cold atomic cloud in MOT by short-distance TOF method for our small vapor cell. If the cold atomic cloud is released, it will fall and expand due to gravity and its initial velocity distribution. Utilizing a cylindrical resonant probe beam positioned several millimeters beneath the MOT ( in this paper, we analyzed the cases for 3mm, 5mm, and 8mm of center-to-center distance between the MOT region and the probe beam, respectively). we can record the short-distance TOF absorption spectrum with a photodiode when cold atoms go through the probe region. Analysis of the short-distance TOF absorption spectrum yields the effective temperature of the cold cloud in MOT.
What problem does this paper attempt to address?