Evaluation the Visualization Quality of Lung Markings on Monochrome LCDs with Different Resolutions

Yin Jiandong,Zhang Wei,Guo Qiyong
DOI: https://doi.org/10.3969/j.issn.1005-5185.2010.02.013
2010-01-01
Abstract:Purpose To evaluate the influence of medical monochrome LCDs with different resolutions on the presentation of lung markings. Materials and Method Three radiologist on high-,mid-and lower-experienced level interpreted 93 chest images on three types of displays with different resolutions. The observation was focused the lung markings of superior-middle field in the right lung. All observers rated their confidence of the visual quality of lung markings with three-point scale:excellent,fine and inferior. Software SPSS 13.0 was used to analyze the interpreting results Result:For the visualization of lung markings,there was no significant difference on different displays used by the same experienced radiologist (P>0.05). There was no significant difference among radiologists with different experience using either 2MP or 3MP display. However,on 5MP display,the difference was significant between lower-and high-experienced radiologists and between mid-and high-experienced radiologists,where t=2.99 and 2.11,respectively,and P<0.05. But there was no significant difference between mid-and lower-experienced radiologists. Conclusion High-experienced radiologist can get more information on high resolution display as 5MP. Mid-and lower-experienced had same performance in interpreting the lung markings on the displays with same middle or lower resolutions.
What problem does this paper attempt to address?