Image Study of TSRH Spinal System for Spinal Fracture

Li Jinzhu,Yang Guangfu
DOI: https://doi.org/10.3969/j.issn.1006-7035.2008.04.001
2008-01-01
Abstract:Purpose:To study the image findings of TSRH spinal system for spinal fracture,and evaluate the clinical treatment.Materials and Methods:63 cases of TSRH spinal system for spinal fracture were obtained.There were single vertebral fracture in 61 cases,and multiple vertebral fracture in 2 cases.Before operation,CR films and CT scans were obtained in all cases,and MRI scans in 12 cases.Results:(1)The successful operation rate was 96.8%;(2)The reposition fineness rate was 95.2% for vertebral fracture collapse;92.4% for fracture fragment;87.0% for spinal dislocation;78.7% for disc colloapse.(3) The position fineness rate for vertebral arch screw was 93.8%.No screw loose,and broken crook and plate were showed in our group.Conclusion:TSRH spinal system is a satisfactory method for spinal fracture,and CR is significance for follow up of TSRH spinal system.
What problem does this paper attempt to address?