Multiparameter comparison research of bipolar radiofrequency ablation on isolated pig hearts

CHEN Ying-min,LI Min,XU Jin,LUO Zhang-yuan,ZHANG Wen-zan,QIN Yong-wen
DOI: https://doi.org/10.7695/zgxzqb201306012
2013-01-01
Abstract:Objective To investigate the ablation lesion size of clinical bipolar radiofrequency ablation on vitro pigs hearts in different conditions and evaluate the effectiveness and safety of bipolar radiofrequency ablation. Methods Isolated pigs hearts( total of 144,each ablative combination of the 6 times) placed in water bath 37 ℃ Langendorff system, with a 4 mm bipolar electrode,electrode distance from 12 mm to 17 mm,time from 20 s to 90 s,a non-saline or cold saline perfusion( 1 000 ml / h),respectively,for ventricular ablation. Our research determinate the connective ablation lesions in length,width,depth to evaluate the effect of connective ablation. Results As time increases,the length,depth,connection rate,blasting rate were increased; as distance increasing,the increasing in length,the depth,connection rate, blasting rate were reduced; When perfused,the length,depth,connection rate,blasting rate were increased. The effectiveness and safety were the best with 12 mm,20 s,1 000 ml / h,the mean length was 22. 17 mm,the average depth was 3. 50 mm,connection rate 100%,blasting rate 0%. Conclusion In the power 30 W and temperature 60 ℃,the bipolar ablation in the electrode with 12 mm,20 s,1000ml / h to form the most stable continuous damage.