Effect of central corneal thickness on Goldmann applanation tonometer

Hong-Bo Yin,Hong-Min Yun,Ya-Li Liu,Man Yu,Xu-Yang Liu,Xiao-Ming Chen
IF: 1.645
2008-01-01
International Journal of Ophthalmology
Abstract:AIM: To study the distribution of central corneal thickness (CCT) in a suburb population and to determine the effect of CCT on the measurement of intraocular pressure (IOP). ·METHODS: In this cross-sectional study, 169 normal subjects were evaluated. CCT was measured with an optical pachymeter by an observer. IOP was measured with Goldmann applanation tonometry by another observer using blind method. ·RESULTS: Mean CCT was 0.547mm (95% confidence intervals 0.543-0.751mm) in right eye and 0.551mm (95% confidence intervals 0.553-0.749mm) in left eye. Mean IOP was 15.8mmHg in right eye and 15.5mmHg in left eye. There was no significant relationship between CCT and age, but a positive relationship was found between CCT and IOP. Linear regression analysis showed an increase of 1mmHg in IOP with each 0.018mm increase in central corneal thickness. ·CONCLUSION: Variation in central corneal thickness is a significant source of variation in IOP measurement. Ocular hypertension can be misdiagnosed by pseudo-hypertension in the patients with thick CCT, and normal tension glaucoma can be misdiagnosed by primary open-angle glaucoma in the patients with thin CCT. Pachymetry should be included in the diagnosis of glaucoma or ocular hypertension, particularly when IOP is inconsistent with other clinic data.