Use Of A Reference Tissue And Blood Vessel To Measure The Arterial Input Function In Dcemri

Xiaobing Fan,Chad R. Haney,Devkumar Mustafi,Cheng Yang,Marta Zamora,Erica J. Markiewicz,Gregory S. Karczmar
DOI: https://doi.org/10.1002/mrm.22551
2010-01-01
Magnetic Resonance in Medicine
Abstract:Accurate measurement of the arterial input function is critical for quantitative evaluation of dynamic contrast enhanced magnetic resonance imaging data. Use of the reference tissue method to derive a local arterial input function avoided large errors associated with direct arterial measurements, but relied on literature values for K-trans and v(e). We demonstrate that accurate values of K-trans and v(e) in a reference tissue can be measured by comparing contrast media concentration in a reference tissue to plasma concentrations measured directly in a local artery after the 1-2 passes of the contrast media bolus when plasma concentration is low and can be measured accurately. The values of K-trans and v(e) calculated for the reference tissue can then be used to derive a more complete arterial input function including the first pass of the contrast bolus. This new approach was demonstrated using dynamic contrast enhanced magnetic resonance imaging data from rodent hind limb. Values obtained for K-trans and ve in muscle, and the shape and amplitude of the derived arterial input function are consistent with published results. Magn Reson Med 64:1821-1826, 2010. (C) 2010 Wiley-Liss, Inc.