Comparison of Three ELISA Kits for Detection of Mycoplasma Bovis Antibody

LIU Yang,CHEN Wei,SONG Zhi-qiang,SUN Wen-jing,LI Yuan,ZOU Xiao-hui,XIN Jiu-qing
DOI: https://doi.org/10.3969/j.issn.1008-0589.2011.06.09
2011-01-01
Abstract:To compare HVRI kit developed by us with the commercial kit 1 and kit 2,thirty-eight positive sera samples(19 sera samples from naturally infected calves and 18 sera samples from experimentally infected calves) and 37 negative sera samples were detected by the 3 ELISA kits.The comprehensive diagnosis results revealed that the detection rates of the HVRI Kit,commercial Kit 1 and Kit 2 were 92%,96% and 74.67%,respectively.The Kappa consistency check of the 3 kits indicated that there was strong consistency between HVRI kit and Kit 1,while there was moderate consistency between Kit 2 and HVRI kit or between Kit 2 and Kit 1.Moreover,test on the international standard positive serum of contagious bovine pleuropneumonia(PS2),both HVRI kit and kit 1 showed negative detection results,but it was positive by kit 2 detection.In conclusion,HVRI Kit and Kit 1 were recommended as reliable method both in routine diagnostics and in epidemiological survey.
What problem does this paper attempt to address?