[Adaptive Removal of Motion Artifact from ECG Based on Impedance Detection].

Shaojie Lin,Lijuan Lai,Xiaoming Wu
2010-01-01
Journal of Biomedical Engineering
Abstract:Motion artifact is a significant source of noise in ambulatory physiologic signals and can affect the diagnoses seriously. To cope with this problem, we adopted certain methods for removing motion artifact from ECG signals via synchronous acquisition of the elector/skin impedance as the reference signal of adaptive filter. The results demonstrated that 3rd order RLS adaptive filter can remove the baseline shift effectively with an exciting current of 120 Hz and 50 microA. Such system was implemented to investigate the utility of ambulatory heart rate monitoring, which provides convenient and reliable monitoring services for patients.