Comparative Investigation of Different Pretreatment Methods for Determining Zn, Fe, Se,Pb in Rice Residue

Sheng-jun Chen,Hua Xiong,Chong-xue Ouyang,Quan-cai Zhang,Liang Li
DOI: https://doi.org/10.3969/j.issn.1004-8685.2007.10.008
2007-01-01
Abstract:Objective: Reference was provided for measuring the levels of microelement. Methods: After the rice residue being pretreated by conventional dry calcination, the wet digestion and microwave digestion, respectively, Zn, Fe, Se, Pb were determined by flame atomic absorption spectrometry (FAAS). Results: The measured values of Zn obtained by the three methods were similar. The measured values of Fe, Pb obtained by the wet or microwave digestions were higher than that obtained by the dry calcination. The measured value of Se obtained by the microwave digestion was the highest among three digestion pretreatments. The precision and the recovery ratio were also different. Precision: the microwave digestion> the wet digestion> the conventional dry calcination. Recovery rate: The recovery rate by the microwave digestion was almost closed to 100%, the recovery rate by the dry calcination was relatively lower, the recovery rate by the wet digestion was slightly higher. Conclusion: Different methods of pretreatment influence results to a certain extent. The microwave digestion is with advantages of high digestion ability and speed, of small loss, less pollution and lower reagent blank, and can be used for the measurement of microelement in the majority.
What problem does this paper attempt to address?