Analysis of Residual Solvents in PET Radiopharmaceuticals by GC

LI Yun-gang,ZHANG Xiao-jun,LIU Jian,TIAN Jia-he,ZHANG Jin-ming
DOI: https://doi.org/10.7538/tws.2013.26.03.0152
2013-01-01
Abstract:The residual solvents in PET radiopharmaceuticals were analyzed by GC, which were acetonitrile, ethanol, N, N-dimethylethanolamine (DMEA), dimethylsulfoxide (DMSO). The standard curves were established with the AT-624 capillary column at GC, and the sensitivity of acetonitrile and ethanol were 0.004-0.320 g/L and 0.010-0.120 g/L respectively. The residual solvents of acetonitrile, ethanol, DMEA and DMSO in PET radio- pharmaceuticals were analyzed by GC. The linearity were 0.9994, 0.9999, 0.9997, 0.999 6 respectively. The residual of acetonitrile were (0.0313±0.0433), (0.0829±0.0668), (0.0156±0.0059), (0.0254±0.0266) g/L in {sup 18}F-FDG, {sup 18}F-FLT, {sup 11}C-CFT, {sup 11}C-PIB respectively. The residual of ethanol was (0.0505±0.00528) g/L in {sup 18}F-FDG. The residual of DMSO were (0.0331±0.0180) g/L, (0.0238±0.0100) g/L in {sup 18}F-W372 and {sup 11}C-DTBZ respectively. The residual of DMEA was (0.0348±0.0022) g/L in {sup 11}C-Choline. The survived of organic solvent in PET radiopharmaceuticals can be analyzed with GC directly. The results showed that the QC should be done in PET radiopharmaceuticals purity with semi-HPLC to avoid the high residual. (authors)
What problem does this paper attempt to address?