Clinical Evaluation of the Gold Immunochromatographic Assay for Quantitative Determination of Β-Hcg

JIANG Hai-yan,DU Min,KE Dong-zhong
DOI: https://doi.org/10.3969/j.issn.1006-1703.2011.01.014
2011-01-01
Abstract:To evaluate the reliability and accuracy of the gold immunochromatographic assay(GICA) for fast quantitatively determination of β-HCG and its clinical value.The β-HCG levels of specimen were tested by GICA and chemiluminescent analyzer and analyzed the coincident rate and correlation of the two methods on the base of clinical diagnosis.The results showed that there was a good correlation between the results of the two methods.The coincident rate of positive samples was 97.9%,and the rate of the negative samples was 96.4%.There was no significant difference between the results measured by two methods(P0.05).The results indicated that GICA may be a simple and effective method in fast quantitative determination of β-HCG and possesses a fixed clinical application value.
What problem does this paper attempt to address?