Determination of organic acids in wei li acid by ion chromotography

YAN Wei,JIAO Xia,YE Ming-li,ZHU Yan
DOI: https://doi.org/10.3969/j.issn.1000-0720.2008.02.010
2008-01-01
Abstract:In this paper,a method for the simultaneous determination of four kinds of organic acids in Wei Li acid is proposed.The organic acids including citric acid,lactic acid,malic acid and fumaric acid were ananlyzed by means of ion chromatography using a gradient elution with heptafluorobutyric acid as mobile phase,Ion Pac ICE-AS6 as the separation column and suppressed conductivity detection. Optimized analytical conditions were further validated in terms of accuracy,precision and total uncertainty.The results showed the reliability of the IC method.Under favorable chromatographic conditions,good linear relationship,sensitivity and reproducibility were obtained.Detection limits of citric acid,lactic acid, malic acid and fumaric acid were 0.10、 0.14、 0.30 and 0.12 μg/L,respectively.Relative standard deviations of repeated analyses were less than 4.3%(n=10).Recoveries were 94%~105%.
What problem does this paper attempt to address?