Techniques of MR Carotid Angiography with a Philips 3.0T Scanner

赵子周,赵辉林,曹烨,戈欣,路青,许建荣
DOI: https://doi.org/10.3969/j.issn.1000-0313.2011.07.027
2011-01-01
Abstract:Objective:To evaluate the capability of bright blood and black blood imaging with a 3.0T MR scanner in the screening of the carotid atherosclerotic lesions.Methods:A phased array coil dedicated to carotid artery was used in 31 stroke patients with suspected arteriosclerosis.Multiple sequences including bright blood 3D time of flight(TOF),black blood T1WI,T2WI sequences were applied.The degree of carotid artery stenosis and the various lesions were classified.The positive rate of bright blood and black blood sequences in detecting lesions were compared.Results:All of the 31 patients had the examination completed.With 3D-TOF,there were normal arterial lumen(n=8),mild stenosis(n=10),moderate stenosis(n=8) and severe stenosis(n=5) detected,the positive rate was 50%,60%,75%,100%,respectively.Conclusion:With a surface coil dedicated for carotid artery,bright blood 3D-TOF sequence together with black blood sequence for demonstrating vascular wall were capable effectively in displaying the atherosclerotic plaque and degree of stenosis of carotid artery.