Bioassay of Glutathione S-transferase (GST) Antibody-antigen Interactions Using Microcantilever Sensor

XUE Chang-guo,LI Kai,ZHU Juan-juan,ZHANG Qing-chuan,LIU Hong,WU Xiao-ping,LIU Jing
DOI: https://doi.org/10.3969/j.issn.1001-4888.2007.03.028
2007-01-01
Abstract:We report on a novel technology of Microcantilever Sensor(MCS) platform for label-free detection of GST (Glutathione-S-transferase) and GST antibody binding. GST-Ag was immobilized on the gold-coated side of the cantilever by using thiol self-assembled monolayer. Consequently the deflections of the microcantilever corresponding to the process of injecting GST-Ab were generated, which was measured by using optical lever technique. The activity of the immobilized antigen and the specificity of the antigen-antibody binding were confirmed by using an Enzyme-Linked Immunosorbent Assay (ELISA). The results show that: a. the MCS can effectively monitor the cantilever deflections corresponding to antigen-antibody binding in real-time; b. the MCS may bring a new method to monitor various interactions between biology macromolecules.
What problem does this paper attempt to address?