A Highly Sensitive PVC Membrane Iodide Electrode Based on Complexes of Mercury(ii) As Neutral Carrier
Y. -Q. Chai,R. Yuan,L. Xu,W. -J. Xu,J. -Y. Dai,F. Jiang
DOI: https://doi.org/10.1007/s00216-004-2772-1
2004-01-01
Analytical and Bioanalytical Chemistry
Abstract:A novel solvent polymeric membrane electrode based on bis(1,3,4-thiadiazole) complexes of Hg(II) is described which has excellent selectivity and sensitivity toward iodide ion. The electrode, containing 1,4-bis(5-methyl-1,3,4-thiadiazole-2-yl-thio)butanemercury(II) [Hg(II)BMTB(NO3)4], has a Nernstian potentiometric response from 2.0 x 10(-8) to 2.0 x 10(-2) mol L(-1) with a detection limit of 8.0 x 10(-9) mol L(-1) and a slope of -59.0+/-0.5 mV/decade in 0.01 mol L(-1) phosphate buffer solution (pH 3.0, 20 degrees C). The selectivity sequence observed is iodide>bromide>thiocyanate>nitrite>nitrate>chloride>perchlorate>acetate>sulfate. The selectivity behavior is discussed in terms of the UV-Vis spectrum, and the process of transfer of iodide across the membrane interface is investigated by use of the AC impedance technique. The electrode was successfully applied to the determination of iodide in Jialing River and Spring in Jinyun Mountains, with satisfactory results.