Online Preconcentration of Recombinant Arg-Gly-Asp-hirudin Using Dynamic Ph Junction for Analysis in Human Urine Samples by Capillary Electrophoresis-Mass Spectrometry.

Shengmin Su,Yunqiu Yu
DOI: https://doi.org/10.1016/j.chroma.2008.12.050
IF: 4.601
2008-01-01
Journal of Chromatography A
Abstract:A sensitive and effort-saving method was established and validated for the quantitative determination of recombinant Arg-Gly-Asp-hirudin (rRGD-hirudin) in human urine samples. The assay was performed on a uncoated fused silica capillary of 70 cm x 50 microm I.D. and a positive voltage of 30 kV was applied. The sample was injected under pressure of 50 mbar for 300 s and the temperature of capillary was kept 25 degrees C. Sheath liquid consisting of 30% methanol and 70% of 0.1% formic acid aqueous solution flowing at 7 microL/min was supplied to the CE-electrospray interface. Utilizing the dynamic pH junction technique, a lower limit of quantitation of approximately 35 nM was achieved (concentration coefficiency was about 100-fold) without complex sample preprocessing procedure. CE-MS conditions and parameters were also optimized to obtain better performance. The method has been successfully applied in clinical research of rRGD-hirudin.
What problem does this paper attempt to address?