Fabrication of a copper nanoparticle/chitosan/carbon nanotube-modified glassy carbon electrode for electrochemical sensing of hydrogen peroxide and glucose

Ying Wang,Wanzhi Wei,Jinxiang Zeng,Xiaoying Liu,Xiandong Zeng
DOI: https://doi.org/10.1007/s00604-007-0844-6
IF: 5.7
2007-01-01
Microchimica Acta
Abstract:. A high-performance amperometric glucose biosensor was developed, based on immobilization of glucose oxidase (GOx) on a copper (Cu) nanoparticles/chitosan (CHIT)/carbon nanotube (CNT)-modified glassy carbon (GC) electrode. The Cu and CNT had a synergistic electrocatalytic effect toward the reduction of hydrogen peroxide in the matrix of biopolymer CHIT. The Cu/CHIT/CNT modified GC electrode could amplify the reduction current of hydrogen peroxide greatly. Besides, the Cu/CHIT/CNT modified GC electrode reduces hydrogen peroxide at a much lower applied potential and inhibit the responses of interferents. With GOx as an enzyme model, a new glucose biosensor was fabricated. The sensitivity of the sensor is due not only to the large microscopic area but also to the high efficiency of transformation of H 2 O 2 generated by enzymatic reaction to current signal. The biosensor exhibited excellent sensitivity (the detection limit is down to 0.02 mM), fast response time (less than 4 sec), wide linear range (from 0.05 to 12 mM), and perfect selectivity.
What problem does this paper attempt to address?