Experimental valuation of Vitreon as intraocular implants

MA Yu,JIANG Yan-Rong,LI Xiao-Xin,YIN Hong
DOI: https://doi.org/10.3969/j.issn.1003-5141.2007.04.008
2007-01-01
Abstract:Objective To evaluate the safety of Vitreon used as intraocular implants in the model of New-Zealand rabbit.Methods Twenty-four rabbits(24 eyes) were random divided into four groups: lensectomy and vitrectomy group(8 eyes),vitrectomy group(8 eyes),control lensectomy and vitrectomy group(4 eyes),control vitrectomy group(4 eyes).The eyes were filled with Vitreon in vitrectomy,the retina and chamber angle were evaluated by light and/or electron microscopy at 1 week,2 weeks,4 weeks,and 8 weeks after operation.Results Histologically chronic inflammatory reactions in the inferior trabecular meshwork and iris stroma,and closure of the chamber angle were observed at 2 months after operation.Vacuoles in the nerve fiber layer in the inferior retina were observed after 1 week.Membrane tissue and macrophages were found after 1 month.Thinning of all layers of retina was observed,specially for the nulear and photoreceptor layer in the inferior retina after 2 months.After 2 weeks after operation,the electron microscopy examination showed that the retinal pigment epithelial cells engulfed the membranous disc of outer segments; The membranous disc of outer segment were broken or disappeared in the inferior retina at 2 months after operation.Conclusion Vitreon is not suitable as the long-term intraocular implants for the damages to the retina.