Evaluation of Flexibility of Three Access Devices for Transumbilical Laparoscopic Surgery by Using Strain Force Measurement System

Xie Xiaofeng,Zhu Jiangfan,Zou Qianlin
DOI: https://doi.org/10.3969/j.issn.1009-6604.2013.02.026
2013-01-01
Abstract:Objective To evaluate the flexibility of three common access devices in transumbilical laparoscopic surgery.Methods A strain force measurement system,which was composed of a training simulator,strain gauges and force sensing systems mounted with a dedicated software,was set up.Based on this system,the forces and operaton time were compared among Multi-ports,TriPort and SILS Port,when a standard task was performed.Results Compared to TriPort group,Multi-ports group had significantly lower mean operation force [right hand:(0.68±0.12) N vs.(1.33±0.22) N,P=0.000;left hand:(0.66±0.18) N vs.(1.25±0.21) N,P=0.000].The SILS Port group showed significantly lower mean operation force than TriPort group [right hand:(0.92±0.29) N vs.(1.33±0.22) N,P=0.003;left hand:(0.82±0.18) N vs.(1.25±0.21) N,P=0.001].And Multi-ports group had significantly shorter operation time than both SILS Port [(76.6±17.8) sec vs.(105.9±27.2) sec,P=0.030) and TriPort groups [(76.6±17.8) sec vs.(106.3±23.9) sec,P=0.028].Conclusion Considering its least force and shortest operation time,we recommend Multi-ports as the most flexible device in a training simulator.