Accuracy of target-controlled infusion of propofol in patients undergoing partial hepatectomy

LI Min,XUE Zhang-gang,JIANG Hao
2010-01-01
Abstract:Objective To evaluate the accuracy of Diprifusor target-controlled infusion(TCI) system for propofol in patients undergoing partial hepatectomy.Methods Twenty-four patients scheduled for selective surgery under combined general and epidural anaesthesia were divided into 2 groups of partial hepatectomy(group H) and conventional upper abdominal surgery(group C) with 12 cases each.Anesthesia was induced with fentanyl 4 μg/kg and TCI propofol 3 μg/ml and maintained with TCI propofol and epidural block.Heart rate(HR),mean arterial pressure(MAP),bispectral index(BIS) and auditory evoked potential index(AAI) were monitored during anesthesia.Arterial blood samples were taken for the measurement of plasma propofol concentrations.The individual prediction error(PE) and pooled bias [median performance error(MDPE)],inaccuracy[median absolute performance error(MDAPE)] and wobble of the propofol TCI device were determined from the pooled and intrasubject performance errors.Results There were no significant differences in HR,MAP,BIS,AAI,PE,MDPE,MDAPE and wobble between two groups.PE,MDPE,MDAPE and wobble of Diprifusor TCI device were 15.4%,11.93%,17.89% and 13.09%,respectively.Conclusion Diprifusor TCI device for propofol may be safely and effectively used in patients undergoing partial hepatectomy under general anesthesia combined with epidural anesthesia.