Determination of Lactofen by High Performance Capillary Electrophoresis

YAN Feng,GUAN Jin,CHEN Xing,SHI Shuang,FU Ye-rong,LIU Xiao-xia
DOI: https://doi.org/10.3969/j.issn.1006-0413.2010.09.011
2010-01-01
Abstract:A high performance capillary electrophoresis (HPCE) method has been established for the analysis of lactofen. The separation was carried out on a fused-silica uncoated capillary column(42 cm×50 μm i.d.). The analysis was completed in less than 6 min in 50 mmol/L borox-150 mmol/L sodium dihydrogen phosphate buffer(pH 9.5) at 15 kV of separation voltage and 281 nm of detection wavelength. A good linearity was obtained for lactofen in the range from 10 to 240 mg/L with correlation coefficient was 0.9997. The limits of detection(LOD) and quantification (LOQ) were 0.8 mg/L(S/N 3) and 2.0 mg/L(S/N 10),respectively. The average recoveries were in the range of 98.2-99.0%,and relative standard deviations(RSDs) were between 0.59 and 2.6%.
What problem does this paper attempt to address?