Time-resolved Post-Chemiluminescence Analysis for Simultaneous Determination of Amikacin and Hydrochlorothiazide

Guo Wei,Yan Rui-Fang,Yuan Jian-Mei,Zhang Zhu-Jun,Miao Jiang-Huan
DOI: https://doi.org/10.3724/sp.j.1096.2012.20255
2013-01-01
Abstract:Based on the differential chemiluminescent rate of amikacin (AMK) and hydrochlorothiazide (HCT) in the system of Luminol-peroxymonosulfate (PMS), a novel time-resolved post-chemiluminescence method was established for the simultaneous determination of AMK and HCT. The HCT system gives the highest chemiluminescence intensity at 1 s, whereas the AMK system gives its most intense chemiluminescence emission at 39. 7 s. The kinetic curve showed two separate chemiluminescence peaks at different time without disturbing each other. Under the optimum conditions described above, the analytical performance was obtained. The linear response for the determination of HCT and AMK was in the range from 8. 0 X 10(-5) to 4. 0 X 10(-3) g/L and from 8. 0 X 10(-4) to 4. 0 X 10(-2) g/L. The relative standard diviation of AMK (4. 0 X 10(-3) g/L) and HCT (4. 0 X 10(-4) g/L) are 3. 7% and 2. 4% (n=11), and the detection limit (3) of AMK and HCT are 2. 0 X 10(-4) g/L and 2. 0 X 10(-5) g/L, respectively.
What problem does this paper attempt to address?