Determination of Trace Rhenium in Geological Samples Using Isotope Dilution-Inductively Coupled Plasma Mass Spectrometry

YANG Sheng-hong,QU Wen-jun,DU An-dao,YANG Gang,CHEN Jiang-feng
DOI: https://doi.org/10.3969/j.issn.0254-5357.2006.02.006
2006-01-01
Abstract:A method for the determination of trace rhenium in geological samples (pyrites and rocks) using isotope dilution-inductively coupled plasma mass spectrometry was developed. Carius tube digestion method was used for sample preparation. The total procedure blanks could be limited in the range of 1. 4 - 10 pg. The method was verified by analyzing the national standard reference molybdenite sample and the result was in good agreement with the certified value (17. 39±0. 32) μg/g. The method has been applied to the determination of trace Re (1.27 ng/g) in rock sample with precision of 3. 89% RSD(n=4, 2s).