The corrosion behavior of three alloys and Ti abutment
Yuyan LIU,Yanmin ZHOU
DOI: https://doi.org/10.3969/j.issn.1003-7632.2006.05.014
2006-01-01
Abstract:Objective To evaluate the corrosion behavior of Gold alloy,Titanium alloy,Ni-Cr alloy and Ti implant abutment in vitro.Methods The electrochemical corrosion behavior of above alloys was examined with potentiadynamic polarization technique,scanning electronic microscopy(SEM),X-ray photoelectron spectroscopy(XPS) and X-ray diffractometer to measure and analyse the Ecorr,Ebd and Anodic polarization curves of the materials in artificial saliva at 37±0.5℃(pH=6.0).Results Corrosion potential of E co_(Ni-Cr),E co_(abutment),E co_(Ti6A14V) and E co_(Au-Pt) were-193.4mV,-27.46mV,-14.7mV,225.6Mv respectively;Transpassive potential of Ebd_(Ni-Cr),Ebd_(abutment), Ebd_(Ti6A14V)and Ebd_(Au-Pt) were 0.90V,1.0V,1.0V,1.2V respectively.Ni-Cr alloy exhibited pitting and general corrosion,the others did not demonstrate distinct corrosive trace.From XPS,the content of Ni,O increased,and the content of Cr,Mo,Al decreased;but the other alloys did not present evident change.The main resultant of Ni-Cr, Titanium alloy and Ti implant abutment were Cr_2O_3,TiO_2,TiO_2 respectively;The resultant of Au-Pt was not detected. Conclusion The results showed that the materials had good corrosion resistance;but Gold (Au-Pt) alloy presented the best corrosion resistance.Corrosion resistance of Ti implant abutment alloy was similar to that of Ti6A14V and was far superior to that of Ni-Cr alloy.It is concluded that Ti6A14V and Gold used as supra-structure in dental implantation should be better than Ni-Cr alloy,and the couple Ti6A14V/Ti implant abutment should be ideal.