Construction of a three dimensional finite element model of the mandibular fixed periodontal splint for the periodontally compromised patients

De-li LI,Lei TANG,Jing-yun HAN,Jian-guo TAN
DOI: https://doi.org/10.3969/j.issn.1009-3761.2008.03.022
2008-01-01
Abstract:Objective: To construct three-dimensional finite element model of mandible with full dentition and periodontal splint for biomechanic researches of fixed prosthesis in the treatment of periodontal disease. Methods: After CT scanning, Mimics 8.1, Geomagic studio 8.0, Imagewell 11.0 and MARC engineering design program were employed to constructe 3-D FE model of the mandible with full dentition and fixed splint. The dependability of the model was testified. Results: The constructed model had a good similarity to the morphology of the three dimensional images. The total number of the nodes was 145493 and the number of the elements was 567790. The stress was mainly concentrated in the apical and cervical area of the periodontal membrane and distributed rationally, which indicated that the finite element model was effective. Conclusion: The 3-D FE model of the mandibular fixed splint constructed by CT scanning technology and digital image transfer in combination with finite element analysis is feasible and effective in resembling the clinical situation and could become a platform in the subsequent biomechanic researches of different periodontal splints, abutments and periodontium.
What problem does this paper attempt to address?