Application of High Performance Liquid Chromatography-Electrochemical Method in Determination of Monoamine Neurotransmitter Metabolites Content in Urine
Zhou Yujie,Wang Qirong,Fang Zilong,Shao Jing,Wen An,Li Juntao,Liu Daijun,Zheng Yi,Qin Lu
DOI: https://doi.org/10.3969/j.issn.1000-6710.2013.04.011
2013-01-01
Abstract:Objective To establish a rapid and accurate technique for determination of vanilmandelic acid(VMA),homovanillic acid(HVA),and 5-hydroxyindoleacetic acid(5-HIAA) in human urine.Methods ISO-3100 high-performance liquid chromatography,BM titanium system pump,WPS 3000 autosampler,WAX-1 anion exchange column(3μm,3 × 150 mm),and 5011A dual channel Coulomb electrochemical detector were used in this study.Mobile phases A and B were prepared with 20 mmol/L phosphate buffered saline solution and acetonitrile at the ratios of 75:20(v/v) and 80:25(v/v),respectively.VMA,HVA,and 5-HIAA contents in adult 24h urine were determined.Results The regression equation for VMA,HVA,and 5-HIAA was y = 55.387x-0.2327(r = 0.9998),y = 55.31x+ 3247(r = 0.9996),and 5-HIAA y = 59.752x-0.1938(r = 0.9987),respectively.The detection limit was 5 ng/ml for VMA,5 ng/ml for HVA,and 7 ng/ml for 5-HIAA,respectively.The coefficient of variation(CV) was 2.7% for VMA,2.4% for HVA,and 1.6% for 5-HIAA,and the inter-assay coefficient of variation(CV) was 3.2% for VMA,4.1% for HVA,and 3.1% for 5-HIAA;and the levels of VMA,HVA,5-HIAA in adult 24h urine were 1.65±0.08 mg/day,2.67±0.24 mg/day,1.79±0.13 mg/day,respectively.All of them were accorded to adult normal levels.Conclusion The technique used in this study demonstrates extremely high specificity and effectiveness for separation of anionic compounds.