Biomechanical research for shoulder instability repairing using cortical bone anchor

LI Guang-hui,LIU Yu-jie,SUN Ming-xue,WANG Ai-yuan,LI Zhong-li,WANG Zhi-gang,ZHU Juan-li
DOI: https://doi.org/10.3969/j.issn.1005-1139.2007.06.014
2007-01-01
Abstract:Objective:To evaluate the biomechanical features of a self-made cortical bone anchor for shoulder instability repair.Methods:Fresh porcine fresh tibial cortical bone was made to ancipital cortical anchors 3.5mm×7.0mm;Ethibond NO.2 non-absorbable suture was pulled through the eyelet for preparation.All specimens were divided to two groups,in control group the FastinRC metal screw was used.An ancipital cortical anchor and a metal screw were embeded into the area between humerus head and greater tubercle separately in the same specimen.The shaft of humerus was fixed on the work desk using clamp,and the end of suture was hanged on the biomechanical testing machine,the biomechanical pull-out testing was performed.Results:The failure modes in cortical bone anchor were eyelet failure in 2 cases,suture failure in 1 case,and anchor pullout in 7 cases.The mean failure load was 181.2N.In FastinRC metal screw group the failure modes were suture failure in 2 cases,and anchor pullout in 8 cases.There was no eyelet failure,the mean failure load was 163.8N.There was no statistical difference between the two groups.The mean data of press test in cortical bone anchor was 890.0±124.3N.Conclusion:The biomechanical property of cortical bone anchor reaches the need of initial capability for shoulder instability repairing.
What problem does this paper attempt to address?