Determination of pentachlorophenol in soil with radioactive tracer as internal standard

Zhou Wen-Qiang,Shan Jun,Guo Hong-Yan,Ji Rong
DOI: https://doi.org/10.13232/j.cnki.jnju.2012.06.004
2012-01-01
Abstract:Pentachlorophenol(PCP)is an endocrine disrupting compound,which has been widely used and caused severe soil contamination.PCP in soil can be extracted with ethyl acetate;however,the determination of PCP by high performance liquid chromatography or gas chromatography may be strongly disturbed by the complexity of soil matrixes and impurities in the soil extract,especially when the PCP concentration in soil is low.Thus,a purification of the extract is needed to remove the impurities.In the present study,according to references,a method for determination of PCP in soil was optimized.A self-prepared silica gel column was used for sample purification,and PCP was eluted with a gradient mixture of ethyl acetate and hexane.Radioactive 14C-labeled PCP was applied as inner standard to localize PCP fractions during the elution and to quantify the recovery of PCP during the sample pretreatment process.After this purification,recovery of PCP was 94.3%±2.9% and PCP in the extract can be easily determined by HPLC with a lowest detected concentration of 6.7 μg/kg dry soil.The 14C-tracer combined to HPLC determination was demonstrated to be a rapid and alternative method for soil PCP analysis to gas chromatography,which requires a tedious purification procedure and 13C-PCP as inner standard.
What problem does this paper attempt to address?