Comparison of Two Digital Imaging Systems in Mammography
何长久,后军民,任胜辉,李燎原,董晓蕾,陆皓,何雅坤,任静,许国辉
DOI: https://doi.org/10.3969/j.issn.1672-3511.2012.04.047
2012-01-01
Abstract:Objective To contrast the differences between digital radiography and computed radiography in mammography so as to explore the advantages of DR.Methods Sample 50 cases respectively from the patient photographed by CR,DR,choice the images of 50 cases respectively photographed by CR,DR and confirmed as breast cancer sufferers by operation and pathology.Then observe their differences in different aspects,such as: imaging quality,radiation dose,condition of photograph(tube voltage and exposure),rate of remake,checking time and the accuracy rate of diagnosis etc.Results The A piece rate, B piece rate,waste piece rate is respectively 90%,8%,2% and 60%,28%,12%,the radiation dose is respectively(1.06±0.39) mGy and(1.93±0.33) mGy,the exposure is respectively (96.85±40.65) mAs and(134.15±35.55) mAs,the rate of remake is respectively 2% and 18%,the checking time is respectively(4.07±0.31) minutes and(8.25±0.54) minutes,the accuracy rate of diagnosing breast cancer is respectively 90%,74%.These differences suggest a statistic meaning(P0.05),while the tube voltage is respectively(27.74±1.39) KV and(27.94±1.56) KV,which has no statistic meaning(P0.05).Conclusion In breast photograph,the application of DR has further increased the accuracy rate of diagnosis and imaging quality,at the same time,reduced the radiation dose,rate of remake,checking time and exposure.