In Vitro and in Vivo Characterization of A Coronary Stent Coated with an Elastic Biodegradable Polymer for the Sustained Release of Paclitaxel

Young Min Shin,Kwang Suk Lim,Ji Yong Jin,Sung In Jeong,Young Moo Lee,Heungsoo Shin,Kyung-Soo Kim
DOI: https://doi.org/10.1007/bf03218654
IF: 2.127
2009-01-01
Macromolecular Research
Abstract:In conclusion, we developed a DES with PLCL and paclitaxel. After expansion with a balloon, there was no evidence of cracks or peelings in the PLCL/paclitaxel coated stent. Paclitaxel was releasedin vitro up to 96% relative to the total loading weight over 50 days in a PLCL concentrationdependent manner and elutedin vivo up to 97.3% for 30 days. From the rabbit iliac artery implantation study, the PLCL/paclitaxel-coated stent successfully reduced the formation of neointimal hyperplasia and thrombosis without causing inflammatory reaction after 90 days of implantation. These results suggested that PLCL/paclitaxel-coated stents may be used as an alternative DES system.
What problem does this paper attempt to address?