Two Steps Cellulose Acetate Membrane Electrophoresis Analysis of Oversulfated Chondroitin Sulfate in Contaminated Heparin

Wang Hao,Yu Guang-Li,Zhao Xia,Hao Cui,Hu Yan-Nan,Yin Xiu-Hong
DOI: https://doi.org/10.3321/j.issn:0253-3820.2009.08.010
2009-01-01
Abstract:Oversulfated chondroitin sulfate(OSCS), a kind of contaminant, which may be associated with allergic response in heparin(HP) injection, was separated from contaminated heparin material by strong ion-exchange chromatography. A two step cellulose acetate membrane electrophoresis was established to detect and analyze the content of OSCS in contaminated heparin. The experimental result showed that, HP and OSCS can be completely separated by 0.05 mol/L barium acetate buffer(pH 5.0) and then 0.15 mol/L zinc acetate buffer(pH 6.3), and the detection limit is 0.1 g/L. Gray value integration showed good linear relationship between the gray value and amount of OSCS (r = 0.9934) with average recovery of 102.1%-106.1% and RSD of 4.1%-6.0%.