Simultaneous determination of nine N-nitrosamine species in water by gas chromatograph-mass spectrometer in electron ionization (EI) mode

LI Ting,XIAN Qi-ming
2013-01-01
Abstract:For the EI mode,there were low responses for N-nitrosamines.Large volume of water samples and concentration factor up to 5000 were adopted and detection conditions of GC/MS in the EI mode were explored for the analysis of nine N-nitrosamine species at ng/L level in water.The N-nitrosamines in water were extracted by coconut charcoal.The calibration curves obtained for nine N-nitrosamine species were in the range of 2-1 000 μg/L(r2=0.997 6-0.999 6).The limit of detection(LOD) for nine N-nitrosamine species spiked at 10 ng/L ranged from 0.8 ng/L to 4.3 ng/L by the SD method and ranged from 0.2 ng/L to 5.0 ng/L by the S/N method.The recoveries of nine N-nitrosamine species spiked at 10 ng/L,50 ng/L and 200 ng/L were 75%-103%,62%-101% and 48%-97%,respectively.The recoveries meet the general requirements 70%-130% of the EPA Method 521,except for NDPhA spiked at 50 ng/L(62%) and 200 ng/L(48%).The precision of the method measured by relative standard deviation(RSD) was less than 12% for all N nitrosamines spiked at 10 ng/L,50 ng/L and 200 ng/L levels.The GC-EI/MS method for the analyses of nine N-nitrosamine species at ng/L level in water was feasible.