Iterative reconstruction for image enhancement and dose reduction in diagnostic cone beam CT imaging.
D. Matenine,M. Schmittbuhl,J. D. De Guise,P. Després,S. Bedwani
DOI: https://doi.org/10.3233/XST-190523
Journal of X-Ray Science and Technology
Abstract:BACKGROUND
Iterative reconstruction is well-established in diagnostic multidetector computed tomography (MDCT) for dose reduction and image quality enhancement. Its application to diagnostic cone beam computed tomography (CBCT) is only emerging and warrants a quantitative evaluation.
METHODS
Several phantoms and a canine head specimen were imaged using a commercially available small-field CBCT scanner. Raw projection data were reconstructed using the Feldkamp-Davis-Kress (FDK) method with different filters, including denoising via total variation (TV) minimization (FDK-TV). Iterative reconstruction was carried out using the TV-regularized ordered subsets convex technique (OSC-TV). Signal-to-noise ratio (SNR), noise power spectrum (NPS) and spatial resolution of images were estimated. Dose levels were measured via the weighted computed tomography dose index, while low-dose image quality degradation was estimated via structural similarity (SSIM).
RESULTS
OSC-TV and FDK-TV were shown to significantly improve image signal-to-noise ratio (SNR) compared to FDK with a standard filter, 5.8 and 4.0 times, respectively. Spatial resolution attained with different algorithms varied moderately across different experiments. For low-dose acquisitions, image quality decreased dramatically for FDK but not for FDK-TV nor OSC-TV. For low-dose canine head images acquired using about 1/5 of the dose compared to a reference image, SSIM dropped to about 0.3 for FDK, while remaining at 0.92 for FDK-TV and 0.96 for OSC-TV.
CONCLUSION
OSC-TV was shown to improve image quality compared to FDK and FDK-TV. Moreover, this iterative approach allowed for significant dose reduction while maintaining image quality.