Effect of Continuous Normal Saline Flush on Patency of Arterial Pressure Monitoring Lines

陈香萍,庄一渝,金奇红,陆琴,黄丽敏
DOI: https://doi.org/10.3969/j.issn.1671-9875.2009.02.003
2009-01-01
Abstract:Objective To study the effect of continuous normal saline flush on patency of radial arterial pressure tube. Method Divide 110 cases of patients needing invasive artery pressure monitoring into observation group (n=55) and control group (n=55) randomly. Patients in observation group receive normal saline flush and patients in control group receive routine heparinized saline flush(1 U/ml). Compare the patency of radial artery pressure tube.Result For all kinds of factors,observation group excludes 4 cases and control group excludes 5 cases. The blocking rate is 11.76% in observation group and 6.00% in control group. There is no significant difference between the two groups (P0.05).Conclusion Continuous normal saline flush can keep the patency of radial arterial pressure tube effectively.
What problem does this paper attempt to address?