Determination of fluoride concentrations in toothpastes by using ion chromatography

He Wei,Lin Huancai,Zhi Qinghui
DOI: https://doi.org/10.3969/j.issn.1001-3733.2007.05.005
2007-01-01
Abstract:Objective:To study the feasibility and precision of using ion chromatographic method to measure fluoride concentrations in toothpastes. Methods:Three kinds of fluoride toothpastes, containing NaF, NaMFP and NaF-NaMFP were measured of fluoride concentrations respectively by using ion chromatography (IC) and fluoride selective electrode (FSE) which is recommended by the Chinese Standard for Toothpaste(GB8372-2001). Precision and recovery rate of the two methods were evaluated. 9 tubes daily used fluoride toothpastes were measured by IC to evaluate the disturbance of other ions in the toothpaste. Results:There was no statistically significant difference in precision degree of the two methods (P>0.05). The recovery rate of F- measured by IC and by FSE had no significant difference (1.00 v.s. 0.99; P>0.05) but the recovery rate of MFP2- by IC and by FSE was statistically different (1.00 v.s. 1.08; P<0.05). One result of IC for NaF-NaMFP toothpaste showed very high recovery rate. Conclusion:IC method can be used to determine the fluoride concentrations in fluoride toothpastes. The result indicates that IC method had advantage in determination of total fluoride of NaMFP toothpastes. But the method needs further study to prevent the disturbance of other ions contained in the toothpaste.