Optimization of DSC Experimental Conditions and Correlation Analysis for Gelatinization Temperature Traits of Rice

Zhang Dapeng,Wu Jianguo,Shi Chunhai
2011-01-01
Abstract:Differential Scanning Calorimetry(DSC) is currently the most accurate method for testing the gelatinization temperature(GT) of rice flour.However,three important factors,including rice flour weight with water volume and the speed of temperature rise in DSC,were differently used in many experiments.Orthogonal design was introduced in present experiment to determine the best conditions for the smallest relative standard deviation(RSD) of TAQ20 DSC.The results indicated that the GT measurements could be influenced by these three important DSC conditions.The combination of 14 μL distilled water with 3.0 mg rice flour and 10 ℃/min speed of temperature rise was the best DSC experimental condition because of the smallest RSD of the GT in each experiment combination.21 indica rice flour samples were used for analyzing GT by this best DSC condition,and their alkali spreading value(ASV) and amylose content(AC) were also determined.The results also showed that the correlation coefficients between GT and ASV or AC were-0.75*** or 0.51* respectively,which meant that GT was significantly negative correlated with ASV,but positive with AC.
What problem does this paper attempt to address?