PAA phantom for use in thermal ablation of high intensity focused ultrasound: Phantom fabrication and acoustic parameters measurements

Bing HU,Li-xin JIANG,Ying HUANG
DOI: https://doi.org/10.3969/j.issn.1000-3630.2006.06.015
2006-01-01
Abstract:Objective An optically transparent phantom used for thermal ablation of HIFU has been developed with acoustic parameters measured. Methods The color and transparency of different concentration phantoms were observed. Properties including density, acoustic speed and acoustic attenuation of PAA gel phantom were measured. Results (1)PAA gel phantom was amber and transparent. The color became heavier and transparency decreased with the egg-white concentration increasing. (2)The density of PAA gel phantom was similar to water, which was from 1.0250 g/cm3 to 1.0617 g/cm3. There was a significant correlation between sound speed and phantom concentration. The acoustic attenuation was from 0.125 dB/cm to 0.329 dB/cm, increased with the phantom density and transducer frequency increasing. Conclusions PAA gel phantom was homogeneous and a good mimicking phantom for HIFU thermal ablation. The experiment results had a satisfied reproducibility.