Double Cameras Wireless Capsule Endoscope System Design

LIU Jin,GU Ying-ke,YIN Zheng,XIE Xiang,LI Guo-lin
DOI: https://doi.org/10.3969/j.issn.1001-9944.2013.08.002
2013-01-01
Abstract:In this paper,in order to lower the miss rate of the wireless capsule endoscope(WCE)for esophagus and large intestine,we have proposed a double cameras WCE system,and we have showed the system architecture and the work flow.In the meantime,we have proposed low power methods based on sleep and wake up mode.We have got an average detection times of 14.35,and an average miss time 6.53 minutes.We have also proposed an image compression algorithm for WCE with a high compression rate and low complexity.Through the experiment the image compression rate can reach 88% peak signal-to-noise ratio(PSNR can amount to 39 dB).In the end,we have showed the realization of the hardware system.The experiment showed that the power consumption of the system is 32.5 μW in sleep mode,21 mW at a 4fps rate,and 163 mW at a 40 fps rate,respectively.